Walk-in Type Drug Stability Test Chamber

Walk-in Type Drug Stability Test Chamber
Product Details

Basic Info

Model NO.:BDSW-8
Specification&technical datas
ModelBDSW-8BDSW-12BDSW-20BDSW-28BDSW-50
Interior dimensions(W×H×Dmm)1800×2100×18002700×2100×18004700×2100×20004500×2100×30006000×2100×4000
Exterior dimensions(W×H×Dmm)2350×2300×20003250×2300×20005250×2300×22005050×2300×32006350×2300×4200
MaterialInteriorSUS304/316 Stainless Steel
ExteriorCoated  Stainless Steel
InsulationFireproof PU+glass fibre
PerformanceTemp range0~65ºC
Humidity range30%~95% RH
Temp fluctuation≤±0.5ºC
Temp deviation±2.0ºC; 
Humidity deviation±3R.H(>75% R.H)±5R.H(≤75% R.H )
Cooling time65ºC↓0ºC ≤80min
Heating timeRt ↑65ºC≤40min
Circulating systemMechanical convection system
Cooling  systemCascade Refrigeration (water or air cooling) + HFC Refrigerant R404a/R23
Heating  system(Balance Temperature Control)P.I.D+P.W.M.+S.S.R.
Humidification system(Balance Humidity Control)P.I.D+P.W.M.+S.S.R.
Humidification modeSteam humidification
DehumidificationAutomatic Data Processing (ADP) critical dew point cooling dehumidification
Water supplyDistilled water only
Sensor
Pt100+electronicsensor
Control systemController IPC based touch panel(OYO/TEMI880) 
Resolution 0.01ºC, 0.1%RH;    
InputPT100 /Rotronic Integrated Temperature&humidity Sensor
View windowMultilayer toughed glass view windows with built-in heaters
Safety protectionLeakage protection,Overload circuit-breaker,Overload protection for compressor&blower,overheating&dry heat protection for humidification system,Over Current Protection, Phase-sequence Protection,over-temperature protection  
Power sourceAC 380V(1±10%)V,50±0.5HZ,3ψ 4W+E
Option3Q Validation and Calibration services 
Test Point40ºC 75%RH, 30ºC 70%RH, 25ºC 60%RH
Operational environment5~35ºC
Note:1.We reserve the right to change specifications without prior notice 


2.Customized sizes and configurations available


Walk-in Drug Stability Test Chamber
Features
1. Unique duel system design ensures perfectly safe, in case one system fails, the other system will work automatically, no danger of anything going wrong.
2. Lab Class: high stability of temperature and humidity with prevention of dew congealment
3. Proper air circulation design, making the highest uniformity of temperature and humidity
4. Provide 3Q report of chamber installation and operation
5. Energy saving: Independent modular design with intelligent control of refrigerant, air flow, heating & cooling humidification
6. System Calibration services is optional

Product Description

Feedback